Application Note

Split Beam Vs Single Beam Optics For All Hach Spectrophotometers Including DR6000, DR3900, DR2800, DR2700

Source: Hach

Split beam optics also known as ratio beam or dual beam, imparts greater accuracy and reproducibility than conventional single beam optics. The absorbance signal in conventional single beam instruments can destabilize even during simple applications thus resulting in data inaccuracy. By contrast, split beam optics compensates and stabilizes any signal fluctuations to increase accuracy and reliability even over prolonged usage such as in time course and kinetic measurements.

The split beam principle involves splitting the beam generated by the light source by a half mirror. One of the split beams passes through the sample and is quantified by a detector whereas the other split beam which is representative of the absorbance signal is measured by an independent detector to obtain a signal reference. The ratio of the values from both detectors is then calculated to detect and compensate for any aberration in the energy of the light source of a temporal change of the optical elements and produce the highly stable photometric values. Therefore, highly stable photometric values are obtainable. This always ensures reliable results even in long-time measurement.
 

access the Application Note!

Get unlimited access to:

Trend and Thought Leadership Articles
Case Studies & White Papers
Extensive Product Database
Members-Only Premium Content
Welcome Back! Please Log In to Continue. X

Enter your credentials below to log in. Not yet a member of Water Online? Subscribe today.

Subscribe to Water Online X

Please enter your email address and create a password to access the full content, Or log in to your account to continue.

or

Subscribe to Water Online